OLYMPUS On-Demand Library
Add to View

Professor. Motohiko Kato

Center for Diagnostic and Therapeutic Endoscopy
Keio University School of Medicine
35 Shinanomachi, Shinjuku-ku, Tokyo, Japan

Cutting and Coagulation with DualKnife J


✔ Disk shaped knife tip enabling hooking and cutting even the situation where the muscle layer stands perpendicularly.

✔ Stable, effective dissection was achieved even for challenging situations such as rich fatty tissue or submucosal hematoma.

✔ Hemostasis and vessel management achieved by coagulating with DualKnife J.

✔ Switching to an approach from the lateral toward the center by pulling back the knife, like dissection with an ITknife can sometimes be effective when the muscle layer is close to the target tissue in a narrow space.

Technical Tips of injection with DualKnife J


✔ Aim the appropriate injection site

  • Avoid injury of the blood vessel

  • Target the loose submucosal layer just above the muscle layer

✔ Strongly contacting the tip of the ceramic is the secret for effective injection with DualKnife J, since injection solution is coming from the tip of the ceramic sheath

✔ Using coagulation current very shortly and pushing the sheath into the tissue for reducing the space between the knife and the submucosal layer enables effective injection.

 

Water Pressure Method


✔ Advantage of Water Pressure Method contributing to safe and effective submucosal dissection.

  • Stable visualization of the submucosal layer from initial submucosal dissection. 

  • Facilitated identification of the submucosal layer, differentiated from vessel, fat tissue, muscle layer.

  • Hooking the tissue edge facilitated by water jet.

✔ DualKnife J demonstrates good cutting capability underwater due to its high electro density.
       Changing electrosurgical unit setting from under air condition is not necessary.

Prof. Kato’s Electrosurgical Generator Setting for Colorectal ESD


*Coagrasper
Single Single-use Electrosurgical Hemostatic Forceps
FD-411QR (Opening with: 4mm)

Disclaimer: Any content or information (“Content”) presented herein is illustrative in nature and does not guarantee or represent specific information, outcomes, or results. Olympus Medical Systems Corp. and its parents, subsidiaries, affiliates, directors, officers, employees, agents, and representatives (collectively “Olympus”) does not represent to or warrant the accuracy or applicability of the Content. Under no circumstances shall Olympus be liable for any costs, expenses, losses, claims, liabilities, or other damages (whether direct, indirect, special, incidental, consequential, or otherwise) that may arise from, or be incurred in connection with, the Content or any use thereof.