OLYMPUS On-Demand Library
Add to View

Coagrasper™ Device

Model FD-411UR

Basic specifications

Distal end

Cup opening width: 4 mm
Retracted — 0.3 mm
Cutting knife diameter: ø0.4 mm (except tip)

Sheath

Max. insertion portion diameter: ø3.1 mm
Working length:

FD-411UR — 2,300 mm

For Safer and More Appropriate Use

Information on this page provided by Olympus

Basic Operation of the Coagrasper™ Device

1 Push the slider to open the cups.

2 Rotate the entire handle to adjust the orientation of the grasping forceps (clockwise rotation is smoother than counterclockwise rotation).

3 Guide the grasping forceps to the target region and grasp it. Lift the grasped tissue gently and supply high-frequency current.

Applicable forceps channel diameters

The Coagrasper™ device for upper gastrointestinal applications (FD-410LR) can be used with a scope that has a channel diameter of 2.8 mm or more. The Coagrasper™ device for colorectal use (FD-411UR) should be used with a scope that has a channel diameter of 3.2 mm or more.

Why the applicable channel diameters are different

The Coagrasper™ device for colorectal use (FD-411UR) has a long working length. When it is inserted into a colonoscope while it is angulated, heavy pressure applied to the proximal end will cause deflection of the sheath. To prevent this, the proximal end of the sheath is coated with a reinforcing sheath, which increases the outer diameter of this part of the Coagrasper™ device.

FD-411UR: to be used with L length colonoscopes only